The Mini Rythmic™ PN+ is an ambulatory pump for Parenteral Nutrition, Antibiotherapy, IVIG, Chemotherapy and other continuous infusion therapies. This portable pump has been designed to:
With infusion rates from 0.1 to 400.0 ml/hr (0.1 to 999.9 mg/hr) and a variety of infusion modes choices (Ramp, Intermittent, Continuous and others) the Mini Rythmic™ PN+ can be used for any high-demand conitnuous infusion treatment such as Parenteral Nutrition, Chemotherapy and Antibiotics Infusion therapies.
The pump has been designed to reduce most of unnecessary potential alarms whilst preserving full safety:
Please visit our administration sets page PN+ Infusion Sets for more information.
The pump supports the connection to the MicrelCare infusion control and monitoring system, through the IP Connect device, that brings significant value to this therapy.
MicrelCare allows for instant remote monitoring of infused volumes over time through the web and shows therapy compliance. Another monitored parameter is the pressure and flow restriction of the Central Venous Catheter, port or PICC line which gives early indications of possible occlusion (subject to availability).
Please click on the Specifications Tab on the right to see the specifications of the Mini Rythmic™ PN+ pump.
| Delivery Modes | Continuous, Volume / Time, Volume / Rate, Auto Ramp, 25 Steps, Intermittent |
| Accuracy | +/- 5% |
| Flow rate | 0.1 to 400ml/h |
| Programming units Concentration | ml/mg/g , 0.1 to 99.9 mg/ml, 0.1 to 999.9 g/ml |
| Volume to infuse | 1 to 9999ml |
| Programmable infusion time | 10min to 99hours |
| Auto ramp, up & down time | 1min to 9h59m |
| Early taper down | Yes |
| Delayed start | 0 to 24h |
| Flat rate titration | Yes |
| Air bubble size (to detect air eliminating filter failure) | 0.1ml; 0.5ml; 2ml; off |
| Automatic restart after occlusion release | Yes |
| KVO | 0.1 to 20.0ml/h |
| Occlusion alarm levels | Low (0.5 bar - 7 PSI), Medium (0.9 bar - 13 PSI), High (1.3 bar - 19 PSI) |
| AC input | 11VDC 500mA |
| Lock level | 3 |
| Historic | 6500 time stamped events |
| Historic graph | Volume / Time |
| Connectivity | Portable printer, PC, MicrelCare |
| Alarms | Occlusions, near end and end of infusion, low and depleted battery, open mechanism, preventive maintenance warning, dose limit, bolus limit, air eliminating filter defect (normal use has no air alarms) |